- The lightest defibrillator (10 lbs) with the most monitoring options available
- Rugged transport unit designed for both out-of-hospital critical care and hospital users
- Very simple operation with dedicated controls for monitor, defib, pacing and treatment summary
- Available with a variety of power management accessories, including stand-alone battery rechargers and an ambulance power mount
- Able to transmit diagnostic quality 12-lead data directly to a hospital PC
- Innovative design to allow custom configurations and future upgrades as needs change
Defibrillator:
- Waveform:Biphasic: Low current, impedance compensated, optimized terminating current, truncated exponential biphasic
- Charge Time: < 10 seconds @ 360 J (typical)
- Charge Indicator: Audible and graphic
- Output: Adult paddles, internal paddles, peds adapter and multipurpose hands-free adapter available
- Synchronizer: Delivers energy within 60 msec
- Disarm: Front panel switch
Monitor/Dispay:
- Input: 12-lead PIC: 3-lead, 12-Lead patient cable, paddles or mutipurpose hands free adapter
- Size: 6” diagonal, non-fade
- Type: Color active matrix TFT display
- Display Resolution: 640 x 480 pixels
- Sweep Speed: 25 mm / sec
- Frequency Response: (User-selectable)
- 0.05 to 40 Hz Limited Diagnostic mode
- 0.05 to 150 Hz Diagnostic mode (automatically sets chart recorder response)
- Common Mode Rejection: 90dB (Complies with AAMI EC13-1992 section 3.2.9.10)
- Heart Rate Meter: 20 to 300 BPM
- ECG Output: 1 V/mV
Recorder:
- Type: High-resolution thermal array
- Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters and treatment summary ACLS events
- Paper Width: 50 mm. Paper Speed: 25 mm/sec
- Delay: 6 seconds
- Treatment Summary: 7 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.)
- Automatically logs into memory the type of event, time and ECG sample
- Tx Summary Log: 60 ECG events or 4000 non-ECG events
Battery:
- High Capacity Type: NiMH 12 volt - Welch Allyn SuperPac™
- Low Battery Indicator: Flashing low battery icon on display
- Multiple Paired Contacts: Ensures quick, error-free insertion and backup reliability
- Recharge Rate: 80% in 7.5 hours, 100% in 9.5 hours (For high capacity batteries)
General:
- Size: 12.5 “ x 12.5” x 5.3”
- Operating: 2.0 hours of continous ECG, SPO2, CO2, and BP monitoring or 120 shocks at 360Joules (typical)
- Temperature*: 0 to 45° C
- Humidity* (NC): 15 to 95% RH
- Vibration*: MIL-STD 810E
- Shock / Drop*: MIL-STD 810E
- Altitude: Up to 15,000 feet
Pulse Oximetry:
- Measurement Range: 31 to 99%
- Heart Rate Range: 30 to 250 BPM ±5% or 5 BPM
- Accuracy: 81-99% = ±2%, 70-80% = 3%, 0-69% unspecified
- Update Period: Every pulse after 4 valid pulses
- Alarm Limits: 60 lower limit, 70 upper limit to 99%
MORE ACCESSORIES FOUND HERE
|