A unique combination of multiparameter monitor and defibrillator, the PIC 40 is ideal for day surgery and other ambulatory care environments.
Features:
- 3 or 5-lead ECG device
- Biphasic with manual defibrillation (2 - 360 joules) or optional AED
- Automated External Defibrillation with escalating energy of 200/300/360 (voice and text prompts)
- Synchronized cardioversion
- Nellcor SpO2
- 6.5” color display
- High Performance Welch Allyn SuperPac™ rechargeable batteries using sophisticated NiMH technology
- Innovative design to allow for future upgrades
- Simple operation with dedicated treatment parameter controls
- Pediatric adapters and power options aquired separately
General:
- Weight: 10.5 lbs (4.76 kg) basic configuration.
- Size: 16.7” x 12.5” x 5.3”
- Operating: Time: 2 hours
- Temperature*: 0 to 45° C
- Humidity* (NC): 15 to 95% RH
- Vibration*: MIL-STD 810E
- Shock / Drop*: MIL-STD 810E
- Altitude: Up to 15,000 feet
Monitor/Display:
- Input: 3-lead or 5-lead patient cable, paddles, or multipurpose hands-free adapter
- Size: 6.4” diagonal, non-fade
- Type: LCD color active matrix
- Display Resolution: 640 x 480 pixels
- Sweep Speed: 25 mm / sec.
- Frequency Response: (User-selectable)
- 2 to 20 Hz Limited mode
- 0.5 to 40 Hz Monitor mode
- 0.05 to 150 Hz Diagnostic mode (automatically sets chart recorder response)
- Heart Rate Meter: 20 to 300 BPM
- Heart Rate Alarms: User-selectable
AED:
- Configurable: Basic & Basic+ modes
- Operation: Continuous analysis of ECG advises shock on VFIB, shockable VTACH
- Energy Sequence: User configurable
- Prompts: Audible voice and visual text prompts
Defibrillator:
- Waveform: Biphasic: Low current, impedance compensated, optimized terminating current, truncated exponential biphasic
- Energy Select:
- External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J
- Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50 J
- Charge Time: < 10 seconds @ 360 J (typical) SuperPak
- Output: Adult paddles, internal paddles, peds adapter and multipurpose hands-free adapter available.
- Synchronizer: Delivers energy within 60 msec
- Disarm: Front panel switch
Recorder:
- Type: High-resolution thermal array
- Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters and treatment summary ACLS events
- Paper Width: 50 mm
- Paper Speed: 25 mm/sec
- Delay: 6 seconds
- Frequency Response: Automatically set to monitor’s frequency response
- Tx Summary Log: 60 ECG events or 300 non-ECG events
Battery:
- High Capacity Type: NiMH 12 volt - Welch Allyn SuperPac
- Low Battery Indicator: Flashing low battery icon on display
- Multiple Paired Contacts: Ensures quick, error-free insertion and backup reliability
- Recharge Rate*: 80% in 7.5 hours, 100% in 9.5 hours (* For high capacity batteries)
Power:
- Mains Voltage: 100-240VAC (auto ranging)
- Mains Frequency: 50/60Hz
Pulse Oximetry:
- Measurement Range: 31 to 99%.
- Heart Rate Range: 30 to 250 BPM ±5% or 5 BPM
- Accuracy: 81-99% = ±2%, 70-80% = 3%, 0-69% unspecified
- Update Period: Every pulse after 4 valid pulses
- Alarm Limits: 60 lower limit, 70 upper limit to 99%
- Probe: Reusable finger or multi-site probe
|